HF-TC-641 Oil Iron Content Tester
This Oil Iron Content Tester accurately measures the concentration of iron particles in lubricating oil, hydraulic oil and transformer oil. It effectively monitors mechanical abrasive wear, supporting equipment condition monitoring and regular oil health analysis in laboratories and on-site applications.
Send your requirements, we will quote for you within 12 hours
The company launches the new-generation HF-TC-641 Series Automatic PQ Ferrous Wear Meter. Integrated with multiple technologies and professional algorithms, the instrument features ultra-high sensitivity and repeatability. It can accurately measure the concentration of ferromagnetic wear debris in various oil samples within seconds. Test results are directly presented as concentration values (mg/L or ppm) for easy reading and analysis. It has been widely used globally in defense, aerospace, marine, power, petrochemical, metallurgy, mining, highway transportation and many other industries.
Instrument Features
- High signal-to-noise ratio. Multiple measures are adopted to eliminate interferences from oil matrix, non-ferromagnetic & non-metallic particles, temperature and other factors. The minimum detection limit reaches 1 ppm, superior to other ferrous content testing instruments. Excellent measurement repeatability enables early detection of abnormal equipment wear.
- Ultra-fast measurement. It generates test data within seconds after sample loading, suitable for high-volume sample testing.
- Simple operation. Operators can master the instrument within minutes without professional training.
- On-site testing available. Compact, small-size and vibration-resistant design supports rapid detection under harsh field conditions.
- Compatible with diverse samples. It analyzes ferrous content in mineral oils, synthetic oils, hydraulic oils, gear oils and lubricating greases.
- Safe & eco-friendly. No sample pretreatment or cleaning solvents required. It causes no health hazards to operators.
- Operation software: Chinese & English interfaces optional, runs on Windows system, with functions of data recording, query, calibration and trend analysis.
Technical Parameters
| Item | Specification |
|---|---|
| Measurement Sensitivity | Minimum 1 mg/L |
| Maximum Concentration Range | 20,000 mg/L |
| Ferrous Debris Size | Greater than 1 μm |
| Sample Volume | 2 ml per test |
| Operating Temperature | 10°C ~ 55°C |
| Power Supply | 220V or built-in battery |
| Dimensions | 90×240×150 mm |
| Weight | 1.8 kg |
Operation Steps for Oil Iron Content Tester
- Prepare oil sample and extract 2ml of sample with a clean pipette.
- Start the tester and wait for the instrument to complete self-check and warm-up.
- Inject the oil sample into the testing cell and close the sample chamber.
- Initiate the measurement program; the instrument automatically detects ferrous particle concentration.
- Read test data displayed on the screen (mg/L or ppm) and save records if needed.
- After testing, clean the testing chamber properly for the next measurement.
Packing List
| No. | Item Description | Specification | Qty | Unit | Remark |
|---|---|---|---|---|---|
| 1 | Main Unit of Oil Iron Content Tester | MT100 | 1 | Set | Automatic PQ ferrous wear meter |
| 2 | Power Adapter | 220V | 1 | Piece | For mains power supply |
| 3 | Built-in Rechargeable Battery | Matching host | 1 | Set | For field detection |
| 4 | Pipette | For 2ml sampling | 1 | Set | For oil sample extraction |
| 5 | Sample Vials | Standard | 10 | Pcs | Containing oil samples |
| 6 | Operation Manual | Chinese & English version | 1 | Copy | Operation and maintenance guide |
| 7 | Calibration Certificate | Factory standard | 1 | Copy | Instrument calibration document |
| 8 | Shockproof Foam & Export Carton | Customized packaging | 1 | Set | Transportation protection |
FAQ for Oil Iron Content Tester
Q1: What is the main function of this tester?
A1: It accurately detects the concentration of ferromagnetic wear particles in mineral oil, hydraulic oil, gear oil and grease to monitor equipment abrasive wear.
Q2: What is the minimum detection limit of the instrument?
A2: The minimum sensitivity reaches 1 mg/L (1 ppm), and the maximum measurable concentration is 20,000 mg/L. It can identify iron debris larger than 1 μm.
Q3: How much sample volume is required for one test?
A3: Only 2 ml oil sample is needed for each measurement.
Q4: How long does one measurement take?
A4: The test finishes within several seconds after sample injection, suitable for mass sample testing.
Q5: Can the instrument work for on-site field inspection?
A5: Yes. It is compact, anti-vibration and supports both 220V power supply and built-in battery, applicable to harsh working environments.
Q6: Is solvent needed for sample pretreatment or cleaning?
A6: No sample pretreatment and no cleaning solvent required. The operation is safe and environmentally friendly.
Q7: Does the supporting software support data analysis?
A7: The Windows-based software has Chinese and English interfaces, supporting data recording, query, calibration and trend analysis.
Q8: What is the working temperature range?
A8: The operating temperature ranges from 10°C to 55°C.