HF-GT-205 Digital Equipotential Bonding Tester
Digital Equipotential Bonding Tester is a portable instrument designed to measure equipotential bonding resistance between metal components. It quickly verifies grounding continuity, eliminates potential difference risks, and ensures safe operation for power stations, workshops and anti-static systems.
Send your requirements, we will quote for you within 12 hours
Model HF-GT-205 Digital Equipotential Bonding Tester, also known as Milliohm Meter or DC Earth Resistance Tester, adopts microprocessor technology and four-wire measurement method to deliver high precision and reliable safety performance.
It is a dedicated instrument for testing the bonding quality between metallic components including equipment casings, lightning arresters, ground beams, building structures, cabinet frames, steel bars, water pipes, windows, guardrails, radiators and assembly line frames. It can also measure the resistance between electrical equipment and grounding grids, transformer DC resistance, as well as contact resistance of switches and socket contacts. Widely applied in power systems, meteorology, computer rooms, oil fields, power distribution networks, railway transmission lines, gas stations, factory grounding grids and lightning protection projects. Additionally, it features DC and AC voltage measurement functions.
The complete set consists of main unit, monitoring software, test leads and communication cable. Equipped with a large LCD screen for intuitive readings, it can store up to 500 groups of measurement data. Resistance measuring range: 0.0001Ω ~ 30kΩ; DC voltage range: 0 ~ 1000V; AC voltage range: 0 ~ 750V. The supporting monitoring software enables real-time monitoring, historical data inquiry, curve display, threshold setting, audible alarm, as well as data retrieval, storage, report generation and printing.
Technical Specifications
| Item | Parameter Details |
|---|---|
| Functions | Equipotential bonding resistance test between metal parts, low resistance test, conductor resistance test between grounding grid electrodes, contact resistance test, etc. |
| DC Resistance Range | 0.0001Ω ~ 30.00kΩ, Accuracy: ±1%rdg ±5dgt |
| Resistance Resolution | 0.0001 Ω |
| Measurement Method | Four-wire testing |
| Test Current | ≥1 A |
| Open-circuit Voltage | ≤7 V |
| Measuring Power | ≤15 W |
| DC Voltage Range | 0.0V ~ 1000V DC, Accuracy: ±1.5%rdg ±5dgt |
| DC Voltage Resolution | 0.1 V |
| AC Voltage Range | 0.0V ~ 750V AC, Accuracy: ±1.5%rdg ±3dgt |
| AC Voltage Resolution | 0.1 V |
| Power Supply | DC 6V 4.5Ah rechargeable lithium battery; Continuous working time ≥ 100 hours |
| Backlight | Adjustable grayscale backlight, suitable for dim environments |
| Display Mode | 4-digit large LCD screen with grayscale backlight |
| Measurement Indication | LED flashing during measurement, countdown display on LCD |
| LCD Size | 108mm × 65mm |
| Overall Dimensions | 277.2mm × 227.5mm × 153mm |
| Test Leads | 2 red 5m leads, 2 black 5m leads |
| Measuring Time | Resistance test: approx. 3 seconds/time; Voltage test: approx. 2 seconds/time |
| USB Interface | Available for software monitoring, data upload, storage and printing |
| Communication Cable | 1 USB communication cable |
| Data Storage | 500 groups; marked “MEM” for storage, “FULL” for full memory |
| Data Query | Historical data query function with “MR” symbol prompt |
| Overrange Display | “OL” symbol for over-limit indication |
| Alarm Function | Audible alarm triggers when readings exceed preset threshold |
| Battery Monitoring | Real-time battery level display; low battery reminder for timely charging |
| Auto Power-off | Automatically shuts down after around 15 minutes of inactivity |
| Power Consumption | Standby: Max 30mA (backlight off) Power-on with backlight: Max 43mA Measuring state: Max 2A (backlight off) |
| Weight | Main unit: 2397g (with battery) Test leads: 850g (including portable lead bag) Instrument case: 1200g |
| Operating Temperature & Humidity | -10℃ ~ 40℃, relative humidity ≤ 80%RH |
| Storage Temperature & Humidity | -20℃ ~ 60℃, relative humidity ≤ 70%RH |
| Overload Protection | Terminals C1-C2, P1-P2 withstand AC 280V for 3 seconds |
| Insulation Resistance | ≥10MΩ (between circuit and enclosure at 500V) |
| Withstand Voltage | AC 3700V rms (between internal circuit and enclosure) |
| Electromagnetic Compatibility | Complies with IEC61010-4-3, radiated radio frequency field ≤1V/m |
| Compliance Standards | IEC61010-1, CAT III 600V, Pollution Degree 2; JJG724-1991 Verification Regulation for Digital DC Ohmmeter; JJG166-1993 Verification Regulation for DC Resistance Meter; DL/T967-2007 Verification Regulation for Loop Resistance Tester & DC Resistance Rapid Tester |
Packing List
| Item No. | Description | Qty | Unit | Remarks |
|---|---|---|---|---|
| 1 | Main Unit of Digital Equipotential Bonding Tester | 1 | Set | Built-in rechargeable battery |
| 2 | Red Test Leads (5m) | 2 | Piece | For resistance & voltage testing |
| 3 | Black Test Leads (5m) | 2 | Piece | |
| 4 | USB Communication Cable | 1 | Piece | For data transmission & firmware upgrade |
| 5 | Power Charger | 1 | Piece | Matched 6V charging adapter |
| 6 | Portable Instrument Carrying Case | 1 | Pcs | Shockproof storage box |
| 7 | User Operation Manual (English + Chinese) | 1 | Copy | Detailed operation & parameter introduction |
| 8 | Factory Calibration Certificate | 1 | Copy | Traceable calibration report |
| 9 | Supporting PC Monitoring Software CD | 1 | Piece | For data recording, curve display & report printing |
| 10 | Accessory Storage Bag for Test Wires | 1 | Pcs | For organizing test leads |
FAQ for Digital Equipotential Bonding Tester
1. What is a digital equipotential bonding tester mainly used for?
It adopts four-wire testing method to measure low bonding resistance between metal components, check grounding continuity, and test DC/AC voltage. Widely used in power stations, computer rooms, lightning protection projects and industrial workshops to eliminate potential voltage hazards.
2. Why adopt the four-wire measurement technology?
Four-wire testing effectively excludes the resistance interference of test leads and contact points, ensuring ultra-high accuracy when measuring micro-resistance, which cannot be achieved by ordinary two-wire multimeters.
3. What resistance range can this instrument measure?
Measuring range: 0.0001Ω ~ 30.00kΩ, suitable for equipotential bonding resistance, grounding connection resistance, contact resistance and transformer DC resistance testing.
4. Can it measure DC and AC voltage simultaneously?
Yes. It supports DC 0~1000V and AC 0~750V voltage measurement, convenient for on-site electrical safety inspection besides resistance testing.
5. How long is the continuous working time?
Equipped with 6V 4.5Ah rechargeable battery, continuous working time exceeds 100 hours; low battery prompt will appear when power is insufficient.
6. What functions does the data storage and supporting software have?
The tester can store 500 groups of measurement data. Via USB cable connected to PC, the software realizes real-time monitoring, curve drawing, historical data inquiry, report generation, storage and printing.
7. Does the device have an alarm function?
Yes. Users can preset threshold values. Once the measured resistance or voltage exceeds the set limit, the instrument will send an audible alarm to remind abnormal grounding conditions.
8. Will the instrument shut down automatically?
It will automatically power off after 15 minutes without any operation to save battery power.
9. What is the applicable working environment temperature?
Operating temperature: -10℃ ~ 40℃, humidity below 80%RH; storage temperature ranges from -20℃ to 60℃, adaptable to most indoor and outdoor power site environments.
10. Is the instrument safe for high voltage field operation?
It complies with IEC61010 CAT III safety standard, built-in overload protection, circuit withstand voltage up to 3700V, effectively preventing high voltage damage to the main unit and operators.
11. How long is the test line? What accessories are equipped?
Comes with two 5m red test leads and two 5m black test leads, long enough for large grounding grid on-site measurement.
12. Is regular calibration required?
Yes. The instrument is delivered with a factory calibration certificate. Periodic calibration is recommended to keep long-term measuring accuracy, complying with domestic power industry verification regulations.
13. What occasions need equipotential bonding testing?
Lightning protection grounding engineering, computer room anti-static grounding, substation metal structure connection, petroleum station grounding system, production line equipment grounding, etc.